Remote control commands
R&S
®
LCX Series
207
User Manual 1179.2260.02 ─ 02
Usage:
Query only
INITiate[:IMMediate]
Starts a new measurement.
In manual trigger mode, the command triggers a single measurement cycle. When
completed, the R&S
LCX waits for the next trigger event.
Example:
See
Example "Using the INITiate command"
Usage:
Event
MEASure:ACCuracy?
Queries the accuracy of the last measurement.
The R&S
LCX returns the accuracy of the impedance (
δ
|Z|) in percent, and the phase
angle (
δΘ
) in degrees.
Usage:
Query only
Manual operation:
See
MEASure:CURRent?
Queries the current value following next in the measurement.
Example:
See
Example "Using the MEASure commands"
Usage:
Query only
MEASure:MODE
<Measurement Mode>
MEASure:MODE?
<Measurement Mode>
Selects whether the R&S
LCX starts and continues a measurement, or starts on initi-
ated trigger events.
Parameters for setting and query:
<Measurement
Mode>
CONTinuous | TRIGgered
CONTinuous
Restarts the measurement automatically after a measurement
cycle has been completed.
TRIGgered
Starts a measurement cycle initiated by a trigger signal.
To delay the measurement start to a certain extent, use com-
mand
Example:
See
Example "Using the MEASure commands"
Manual operation:
See
Measurement commands