Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
209
User Manual 1173.9240.02 ─ 23
Example:
SYST:COMM:RDEV:OSC:VDEV?
Usage:
Query only
SYSTem:COMMunicate:RDEVice:OSCilloscope:VFIRmware?
Queries whether the firmware on the connected oscilloscope is supported by the
2
GHz bandwidth extension (R&S
FSW-B2000) option.
Return values:
<State>
ON | 1
Firmware is supported
OFF | 0
Firmware is not supported
Example:
SYST:COMM:RDEV:OSC:VFIR?
Usage:
Query only
TRIGger[:SEQuence]:OSCilloscope:COUPling
<CoupType>
Configures the coupling of the external trigger to the oscilloscope.
Parameters:
<CoupType>
Coupling type
DC
Direct connection with 50
Ω
termination, passes both DC and
AC components of the trigger signal.
CDLimit
Direct connection with 1
M
Ω
termination, passes both DC and
AC components of the trigger signal.
AC
Connection through capacitor, removes unwanted DC and very
low-frequency components.
*RST:
DC
Manual operation:
See
11.4.2.4
Configuring Input via the Optional Analog Baseband Interface
The following commands are required to control the optional Analog Baseband Inter-
face in a remote environment. They are only available if this option is installed.
For more information on the Analog Baseband Interface see the R&S
FSW I/Q Ana-
lyzer User Manual.
Useful commands for Analog Baseband data described elsewhere:
●
INP:SEL AIQ
●
Configuring the Measurement