Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
191
User Manual 1173.9240.02 ─ 23
Note the special conditions and restrictions for the YIG-preselector described in
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1 (0 for I/Q Analyzer, GSM, VSA, Pulse, Amplifier,
Transient Analysis, DOCSIS and MC Group Delay
measurements)
Example:
INP:FILT:YIG OFF
Deactivates the YIG-preselector.
Manual operation:
See
INPut:IMPedance
<Impedance>
This command selects the nominal input impedance of the RF input. In some applica-
tions, only 50
Ω
are supported.
The command is not available for measurements with the optional Digital Baseband
Interface.
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Example:
INP:IMP 75
Manual operation:
See
See
INPut:SELect
<Source>
This command selects the signal source for measurements, i.e. it defines which con-
nector is used to input data to the R&S
FSW.
If no additional input options are installed, only RF input is supported.
Parameters:
<Source>
RF
Radio Frequency ("RF INPUT" connector)
DIQ
Digital IQ data (only available with optional Digital Baseband
Interface
For details on I/Q input see the R&S
FSW I/Q Analyzer User
Manual.
AIQ
Analog Baseband signal (only available with optional Analog
Baseband Interface R&S
FSW-B71)
For details on Analog Baseband input see the R&S
FSW I/Q
Analyzer User Manual.
*RST:
RF
Configuring the Measurement