Configuration
R&S
®
FSW-K7
107
User Manual 1173.9240.02 ─ 23
Resolution Bandwidth
Defines the resolution bandwidth for data acquisition. The available range is specified
in the data sheet.
Remote command:
[SENSe:]BANDwidth[:RESolution]
5.5.2 Sweep Settings
Access
: "Overview" > "Data Acquisition" > "Sweep" tab
The sweep settings define how often data from the input signal is acquired and then
demodulated.
.................................................................................. 107
...................................................................................... 108
...............................................................................................108
.................................................................................... 108
.......................................................................................... 109
.............................................................................................. 109
Continuous Sweep/RUN CONT
After triggering, starts the sweep and repeats it continuously until stopped. This is the
default setting.
While the measurement is running, the "Continuous Sweep" softkey and the RUN
CONT key are highlighted. The running measurement can be aborted by selecting the
highlighted softkey or key again. The results are not deleted until a new measurement
is started.
Note:
Sequencer. If the Sequencer is active, the "Continuous Sweep" softkey only
controls the sweep mode for the currently selected channel. However, the sweep mode
only takes effect the next time the Sequencer activates that channel, and only for a
channel-defined sequence. In this case, a channel in continuous sweep mode is swept
repeatedly.
If the Sequencer is active in MSRT mode, the "Continuous Sweep" function does not
start data capturing. It merely affects trace averaging over multiple sequences. In this
case, trace averaging is performed.
Data Acquisition