Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
255
User Manual 1173.9240.02 ─ 23
[SENSe:]BANDwidth:DEMod:TYPE
<FilterType>
[SENSe:]BWIDth:DEMod:TYPE
<FilterType>
This command defines the type of demodulation filter to be used.
This command is identical to
SENS:ADEM:BAND:DEM:TYPE
:
Parameters:
<FilterType>
FLAT
Standard flat demodulation filter
GAUSs
Gaussian filter for optimized settling behavior
*RST:
FLAT
Manual operation:
See
[SENSe:]BANDwidth[:RESolution]
<Bandwidth>
This command defines the resolution bandwidth and decouples the resolution band-
width from the span.
For statistics measurements, this command defines the
demodulation
bandwidth.
Parameters:
<Bandwidth>
refer to data sheet
*RST:
RBW: AUTO is set to ON; DBW: 3MHz
Example:
BAND 1 MHz
Sets the resolution bandwidth to 1 MHz
Manual operation:
See
[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of sweeps that the application uses to average
traces.
In case of continuous sweep mode, the application calculates the moving average over
the average count.
In case of single sweep mode, the application stops the measurement and calculates
the average after the average count has been reached.
Suffix:
<n>
.
Parameters:
<SweepCount>
When you set a sweep count of 0 or 1, the R&S
FSW performs
one single sweep in single sweep mode.
In continuous sweep mode, if the sweep count is set to 0, a
moving average over 10 sweeps is performed.
Range:
0 to 200000
*RST:
0