Configuration
R&S
®
FSV/A3000 I/Q
Analyzer
82
User Manual 1178.8536.02 ─ 01
Remote command:
5.7.2
Sweep Settings
Access
: "Overview" > "Bandwidth" > "Sweep" tab
................................................................................................. 83
.....................................................................................83
...........................................................................................83
................................................................................................84
Sweep Points
In the I/Q Analyzer application, a specific frequency bandwidth is swept for a specified
measurement time. During this time, a defined number of samples (= "Record
Length" ) are captured. These samples are then evaluated by the applications. There-
fore, in this case the number of sweep points does not define the amount of data to be
acquired, but rather the number of trace points that are evaluated and displayed in the
result diagrams.
Note:
As opposed to previous versions of the I/Q Analyzer, the sweep settings are now
window-specific.
For some result displays, the sweep points may not be editable as they are determined
automatically, or restrictions may apply. For the I/Q vector result display, the number of
I/Q samples to record ( "Record Length" ) must be identical to the number of trace
points to be displayed ("Sweep Points"). Thus, the sweep points are not editable for
this result display. If the "Record Length" is edited, the sweep points are adapted auto-
matically. For record lengths outside the valid range of sweep points, i.e. less than 101
points or more than 100001 points, the diagram does not show valid results.
Using fewer than 4096 sweep points with a detector other than "Auto Peak" may lead
to wrong level results. For details see
"Combining results - trace detector"
Data Acquisition and Bandwidth Settings