Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
241
User Manual 1178.8536.02 ─ 01
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:X:SLIM:ZOOM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation:
See
CALCulate<n>:THReshold
<Level>
This command defines a threshold level for the marker peak search (for
all
markers in
all
windows).
Suffix:
<n>
.
irrelevant
Parameters:
<Level>
Numeric value. The value range and unit are variable.
*RST:
-120 dBm
Default unit: DBM
Example:
CALC:THR -82DBM
Sets the threshold value to -82 dBm.
Manual operation:
See
CALCulate<n>:THReshold:STATe
<State>
This command turns a threshold for the marker peak search on and off (for
all
markers
in
all
windows).
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:THR:STAT ON
Switches on the threshold line.
Manual operation:
See
" Deactivating All Search Limits "
I/Q Analysis