Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
237
User Manual 1178.8536.02 ─ 01
Suffix:
<n>
.
irrelevant
Parameters:
<DisplayMode>
ON | 1
Displays the marker information in the diagrams.
OFF | 0
Hides the marker information in the diagrams.
*RST:
1
Example:
DISP:MINF OFF
Hides the marker information.
Manual operation:
See
CALCulate<n>:MARKer<m>:X:SSIZe
<StepSize>
This command selects the marker step size mode for
all
markers in
all
windows.
It therefore takes effect in manual operation only.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<StepSize>
STANdard
the marker moves from one pixel to the next
POINts
the marker moves from one sweep point to the next
*RST:
POINts
Example:
CALC:MARK:X:SSIZ STAN
Sets the marker step size to one pixel.
Manual operation:
See
8.7.3.3
Configuring and Performing a Marker Search
The following commands control the marker search.
CALCulate<n>:MARKer<m>:LOEXclude
.......................................................................... 238
CALCulate<n>:MARKer<m>:PEXCursion
.........................................................................238
.............................................................................. 238
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
.................................................................239
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
....................................................................239
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
...................................................................240
CALCulate<n>:MARKer<m>:X:SLIMits:ZOOM[:STATe]
...................................................... 240
..............................................................................................241
....................................................................................241
I/Q Analysis