Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
240
User Manual 1178.8536.02 ─ 01
Parameters:
<SearchLimit>
The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST:
left diagram border
Default unit: HZ
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Manual operation:
See
" Search Limits ( Left / Right )"
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
<SearchLimit>
This command defines the right limit of the marker search range for
all
markers in
all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<Limit>
The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST:
right diagram border
Default unit: HZ
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation:
See
" Search Limits ( Left / Right )"
CALCulate<n>:MARKer<m>:X:SLIMits:ZOOM[:STATe]
<State>
This command adjusts the marker search range to the zoom area for
all
markers in
all
windows.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
I/Q Analysis