Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
75
Operating Manual 1176.7610.02 ─ 02
Example:
For the first steps refer to
chapter 4.14, "Programming Exam-
CONF:BTO:MEAS CFST
Activates the Carrier Frequency Stability and Modulation Accu-
racy measurement.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:CFST:FERR:INIT? MAX
Reads the combined maximum frequency deviation of all pack-
ets and all blocks.
Usage:
Query only
Mode:
BT
Manual operation:
See
CALCulate<n>:BTOoth:DPENcoding:[TOTal]?
This command reads the number of tested packets in the Differential Phase Encoding
measurement.
Note:
This command is only available with active Differential Phase Encoding measurement
(see
on page 96). With all other measure-
ments this command causes a query error.
Suffix:
<n>
.
irrelevant
Example:
For the first steps refer to
chapter 4.14, "Programming Exam-
CONF:BTO:MEAS DPEN
Activates the Differential Phase Encoding measurement.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:DPEN?
Reads the number of tested packets.
Usage:
Query only
Mode:
BT
Manual operation:
See
CALCulate<n>:BTOoth:DPENcoding:BER?
This command reads the bit error rate (BER) in the Differential Phase Encoding mea-
surement.
Note:
CALCulate subsystem