Instrument Functions for Bluetooth Measurements (K8)
R&S
®
FSV-K8
48
Operating Manual 1176.7610.02 ─ 02
Principally, setting the channel number has the same effect as changing the center fre-
quency. The major difference is that the center frequency is not limited to available fre-
quency band values, i.e. values outside the frequency band and between the discrete
channels can be selected (see also
softkey in the "Frequency" menu).
Remote command:
Meas Setup ← Carr Freq Stability
Opens a dialog box for common measurement settings, see
Find Sync ← Carr Freq Stability
Opens a dialog box to set the signal processing functions of the analyzer, see
Block Count (Carr Freq Stability) ← Carr Freq Stability
Opens an edit dialog box to enter the number of blocks to be measured. Every block
has the length of 50
μ
s. The default value is 200 blocks.
This softkey is only available in single sweep operation.
Remote command:
Diff Phase
Opens a submenu to configure the Diff Phase measurement. For further details refer to
chapter 3.1.9, "Differential Phase Encoding (EDR) measurement"
Remote command:
CONF:BTO:MEAS DPEN
, see
Channel ← Diff Phase
Opens an edit dialog box to enter the transmission channel number. From the number
of the channel the center frequency is calculated in accordance to the RF Test Specifi-
cation. The default setting is channel number 0.
Principally, setting the channel number has the same effect as changing the center fre-
quency. The major difference is that the center frequency is not limited to available fre-
quency band values, i.e. values outside the frequency band and between the discrete
channels can be selected (see also
softkey in the "Frequency" menu).
Remote command:
Meas Setup ← Diff Phase
Opens a dialog box for common measurement settings, see
Find Sync ← Diff Phase
Opens a dialog box to set the signal processing functions of the analyzer, see
Softkeys of the Bluetooth Menu