Example:
ADEM:AM AVER,MAXH,MINH,OFF,OFF,OFF
Determines average, max hold and min hold values simultane-
ously for the traces 1-3.
ADEM:AM WRIT,OFF,OFF,OFF,OFF,OFF
Determines only the current measurement values for trace 1.
ADEM:AM OFF,OFF,OFF,OFF,OFF,OFF
Switches AM demodulation off.
Mode:
ADEMOD
[SENSe:]ADEMod:AM:RELative[:TDOMain][:TYPE]
<TraceMode>
This command selects the result types to be measured simultaneously by AM demodu-
lation.
Parameters for setting and query:
<TraceMode>
<TraceMode1>, <TraceMode2>, <TraceMode3>, <Trace-
Mode4>, <TraceMode5>, <TraceMode6>
WRITe | AVERage | MAXHold | MINHold | VIEW | OFF
For details on trace modes see
*RST:
WRITe,OFF,OFF,OFF,OFF,OFF (FM-
Stereo:OFF,OFF,OFF,OFF,OFF,OFF)
Example:
ADEM:AM:REL AVER,MAXH,MINH
Determines average, max hold and min hold values simultane-
ously.
ADEM:AM:REL WRIT,OFF,OFF
Determines only the current measurement values.
ADEM:AM:REL OFF,OFF,OFF
Switches AM demodulation off.
Mode:
ADEMOD
[SENSe:]ADEMod:AM:RELative[:TDOMain]:RESult?
<TraceMode>
This command reads the result data obtained by AM demodulation for the specified
result type. The data format of the output data block is defined by the FORMat com-
mand (see
Chapter 2.3.6.2, "Formats for Returned Values: ASCII Format and Binary
The output unit is %.
Query parameters:
<TraceMode>
WRITe | AVERage | MAXHold | MINHold | VIEW
The specified trace mode must be one of those configured by
[SENSe:]ADEMod:AM[:ABSolute][:TDOMain][:TYPE]
on page 125. Otherwise a query error is generated.
For details on trace modes see
Remote Commands of the Analog Demodulation (R&S