Remote Control
R&S
®
FSPN
507
User Manual 1179.4363.02 ─ 01
NEW | FSL
R&S
FSPN format
Format is also compatible to the R&S FSL and R&S FSV family
*RST:
not reset!
Example:
SYST:FORM:IDEN LEG
Adapts the return value of *IDN? to the R&S FSP/FSU/FSQ fam-
ily.
Manual operation:
See
SYSTem:PRESet:FILTer
<FilterType>
This command selects the resolution filter type that is selected after a preset in the
Spectrum application.
Parameters:
<FilterType>
NORMal
Selects 3 dB filter.
NOISe
Selects 3 dB filter.
(NORMal and NOISe have the same effect.)
PULSe
Selects 6 dB filter.
*RST:
NORMal
Example:
//Select the 6 dB filters as the default filter type
SYST:PRES:FILT PULS
11.9.8
Using Service Functions
......................................................................................507
DIAGnostic:SERVice:SFUNction:LASTresult?
................................................................... 508
DIAGnostic:SERVice:SFUNction:RESults:DELete
............................................................. 508
DIAGnostic:SERVice:SFUNction:RESults:SAVE
................................................................508
.......................................................................................... 508
............................................................................................. 509
........................................................................................509
DIAGnostic:SERVice:SFUNction
<ServiceFunction>
This command starts a service function.
The service functions are available after you have entered the level 1 or level 2 system
password.
Parameters for setting and query:
<ServiceFunction>
String containing the ID of the service function.
The ID of the service function is made up out of five numbers,
separated by a point.
Configuring the R&S
FSPN