Remote Control
R&S
®
FSPN
344
User Manual 1179.4363.02 ─ 01
Parameters:
<Voltage>
<numeric value>
Range:
-10 to 28
*RST:
0
Default unit: V
Example:
//Define a minimum voltage of -5 V
SOUR:VOLT:CONT:LEV:LIM:LOW -5
Manual operation:
See
SOURce:VOLTage:CONTrol<i>:LEVel[:STATe]
<State>
This command turns the tuning voltage source (V
tune
) on and off.
Note that DC power is actually supplied only if you additionally activate the outputs in
general.
Prerequisites for this command
●
).
Suffix:
<i>
.
1..n
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
//Turn on the V
tune
voltage source
SOUR:VOLT:CONT:LEV ON
Manual operation:
See
"Turning the output on and off"
SOURce:VOLTage:POWer<i>:LEVel:AMPLitude
<Voltage/Current>
This command defines the output voltage for the V
supply
source.
Suffix:
<i>
.
1..n
irrelevant
Common Measurement Settings