Remote Commands
R&S
®
FSMR3000
557
User Manual 1179.0116.02 ─ 02
The command works on
all
limit lines in
all
measurement windows at the same time.
Suffix:
<n>
.
<li>
irrelevant
Example:
CALC:LIM:CLE
Deletes the result of the limit check.
CALCulate<n>:LIMit<li>:FAIL?
This command queries the result of a limit check in the specified window.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single measurement mode.
See also
Suffix:
<n>
.
<li>
Return values:
<Result>
0
PASS
1
FAIL
Example:
INIT;*WAI
Starts a new sweep and waits for its end.
CALC2:LIM3:FAIL?
Queries the result of the check for limit line 3 in window 2.
Usage:
Query only
12.6.6.4
Programming Example: Using Limit Lines
The following examples demonstrate how to work with limit lines in a remote environ-
ment.
Example: Configuring Limit Lines
..........................................................................557
Example: Performing a Limit Check
......................................................................559
Example: Configuring Limit Lines
This example demonstrates how to configure 2 limit lines - an upper and a lower limit -
for a measurement in a remote environment.
//------------- Configuing the limit lines ---------------------
CALC:LIM1:NAME 'FM1'
//Names limit line 1 'FM1'.
Analyzing Measurements (Basics)