Remote Commands
R&S
®
FSMR3000
542
User Manual 1179.0116.02 ─ 02
ON | 1
Switches the function on
Example:
DEM:SQU ON
Signals below the level threshold are not sent to the audio out-
put.
12.6.5
Configuring Display Lines
The commands required to configure display lines in a remote environment are descri-
bed here.
...............................................................................................542
.................................................................................... 542
............................................................................................... 543
.....................................................................................543
............................................................................................... 544
.....................................................................................544
CALCulate<n>:DLINe<dl>
<Position>
This command defines the (horizontal) position of a display line.
Suffix:
<n>
.
<dl>
1 | 2
Parameters:
<Position>
The value range is variable.
You can use any unit you want, the R&S
FSMR3 then converts
the unit to the currently selected unit. If you omit a unit, the
R&S
FSMR3 uses the currently selected unit.
*RST:
(state is OFF)
Default unit: DBM
Example:
CALC:DLIN2 -20dBm
Positions the second display line at -20 dBm.
Manual operation:
See
"Horizontal Line 1/ Horizontal Line 2"
CALCulate<n>:DLINe<dl>:STATe
<State>
This command turns a display line on and off
Suffix:
<n>
.
<dl>
1 | 2
Parameters:
<State>
ON | OFF | 0 | 1
Analyzing Measurements (Basics)