[SENSe:]AVERage<n>:COUNt
<AverageCount>
This command defines the number of measurements that the application uses to aver-
age traces.
In case of continuous sweep mode, the application calculates the moving average over
the average count.
In case of single sweep mode, the application stops the measurement and calculates
the average after the average count has been reached.
Suffix:
<n>
.
irrelevant
[SENSe:]AVERage<n>[:STATe<t>]
<State>
This command turns averaging for a particular trace in a particular window on and off.
Suffix:
<n>
.
<t>
Parameters:
<State>
ON | OFF | 1 | 0
[SENSe:]AVERage<n>:TYPE
<Mode>
This command selects the trace averaging mode.
Suffix:
<n>
.
1..n
Parameters:
<Mode>
LOGarithmic
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
Example:
AVER:TYPE LIN
Switches to linear average calculation.
Manual operation:
See
Analyzing Measurements (Basics)