Remote Commands
R&S
®
FSMR3000
415
User Manual 1179.0116.02 ─ 02
12.5.1.3
External Generator Control
For each measurement channel one external generator can be configured. To switch
between different configurations define multiple measurement channels.
..................................................................................415
..........................................................................................418
.................................................................................................420
Programming Example for External Generator Control
........................................ 423
Measurement Configuration
The following commands are required to activate external generator control and to con-
figure a calibration measurement with an external tracking generator.
SOURce<si>:EXTernal<gen>:FREQuency
........................................................................415
SOURce<si>:EXTernal<gen>:FREQuency:COUPling[:STATe]
............................................ 415
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:DENominator
......................................416
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:NUMerator
......................................... 416
SOURce<si>:EXTernal<gen>:FREQuency:OFFSet
........................................................... 417
SOURce<si>:EXTernal<gen>:POWer[:LEVel]
................................................................... 417
SOURce<si>:EXTernal<gen>[:STATe]
.............................................................................. 418
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
.............................................................418
SOURce<si>:EXTernal<gen>:FREQuency
<Frequency>
This command defines a fixed source frequency for the external generator.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Frequency>
Source frequency of the external generator.
*RST:
1100050000
Default unit: HZ
Example:
//Define frequency of the generator
SOUR:EXT:FREQ 10MHz
Manual operation:
See
SOURce<si>:EXTernal<gen>:FREQuency:COUPling[:STATe]
<State>
This command couples the frequency of the external generator output to the
R&S
FSMR3.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<State>
ON | OFF | 0 | 1
Configuring the Measurement