................................................................................ 308
User-Defined Parameter Coupling
........................................................................310
...........................................................................314
Example for a User-Defined Parameter Coupling
.................................................315
10.6.1
General Parameter Coupling
Access
: [SETUP] > "Parameter Coupling" > "General"
The "General" tab of the coupling manager contains several parameters that you can
couple across all (active) measurement channels - if the channel supports the corre-
sponding parameter.
When you couple a parameter across all active measurement channels, a change in
the currently selected application is passed on to all other active measurement chan-
nels.
Synchronizing parameters across all measurement channels
....................................308
Selecting all or no coupling mechanisms
.................................................................... 310
Restoring the default configuration
............................................................................. 310
Synchronizing parameters across all measurement channels
To synchronize a specific parameter, turn on the corresponding function.
Note that you cannot synchronize all parameters at the same time, because some
parameters are interdependent. For example, you cannot synchronize the resolution
and video bandwidth simultaneously, because the video bandwidth depends on the
resolution bandwidth and vice versa.
"Center Frequency"
Synchronizes the center frequency
Remote command: