Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
74
User Manual 1179.0880.02 ─ 03
Example:
//Supply 5-pin probe connector with power
OUTP:PROB2 ON
Manual operation:
See
8.5.3
Amplitude configuration
Commands to configure the amplitude described elsewhere.
●
●
............................................................................................. 74
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RLEVel
...............................74
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
........................................................................................... 75
INPut<ip>:ATTenuation:PROTection[:STATe]
.......................................................................76
.........................................................................................76
.............................................................................................. 77
..............................................................................................77
CALCulate<n>:UNIT:POWer
<Unit>
This command selects the unit of the y-axis.
The unit applies to all power-based measurement windows with absolute values.
Suffix:
<n>
.
irrelevant
Parameters:
<Unit>
DBM | V | A | W | DBPW | WATT | DBUV | DBMV | VOLT |
DBUA | AMPere | DBM_mhz | DBM_hz | DBUa_mhz |
DBUV_mhz | DBmV_mhz | DBpW_mhz
*RST:
dBm
Example:
CALC:UNIT:POW DBM
Sets the power unit to dBm.
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RLEVel
<ReferenceLevel>
This command defines the reference level (for all traces in all windows).
With a reference level offset
≠
0, the value range of the reference level is modified by
the offset.
Suffix:
<n>
.
irrelevant
Configuration