Status displays
The status displays on the left side of the diagram indicate any irregularities (e.g. UNCAL).
UNCAL
Indicates that one of the following conditions is present:
• Correction data has been switched off (
CAL
menu,
CAL CORR OFF
).
– Switch on the correction
CAL CORR ON
or
PRESET
.
• No valid correction values are available. This occurs, for example, if a firmware
update is performed followed by a cold start of the instrument.
– Record the correction data (
CAL
menu,
CAL TOTAL
).
OVLD
Indicates an overload of the input mixer.
– Increase the RF attenuation.
IFOVL
Indicates an overload of the IF signal path after the input mixer.
– Increase the reference level.
LOUNL
Indicates that an error was detected in the instrument's frequency processing
hardware.
EXREF
Is displayed if the instrument was set to an external reference but no signal was
detected on the reference input.
OVEN
Indicates that the OCXO reference frequency (option R&S FSU-B4) has not yet
reached its operating temperature. The message usually disappears a few minutes
after power-on.
OVTRC
Indicates a temporary overload condition of the input mixer or IF signal path while
the trace is measured with Max Hold, Min Hold or Average (Overload Trace). OVLD
and IFOVL override this indication.
Summary of Contents for esu
Page 26: ...R S ESU Contents 6 Quick Start Guide 1302 6163 62 04 ...
Page 28: ......
Page 32: ...R S ESU 0 4 Quick Start Guide 1302 6163 62 04 ...
Page 88: ...R S ESU Preparing for Use Operating System Properties 2 38 Quick Start Guide 1302 6163 62 04 ...
Page 104: ...R S ESU Basic Operation Setting Parameters 4 12 Quick Start Guide 1302 6163 62 04 ...
Page 201: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix Appendix ...
Page 202: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix ...