Basic Measurement Examples
Performing a Level and Frequency Measurement
5.14
Quick Start Guide 1302.6163.62 - 04
12.Store the test results, tables or diagrams.
In the example, test results are stored on a memory stick. Measured data are output in
WMF format for use in other applications.
DEVICE1 and output in WMF format is automatically selected with
PRESET
.
➢
Press the
HCOPY
key.
The
HCOPY SETTINGS
menu is displayed where the output of measurement and
device data can be configured and started.
➢
Connect a memory stick to the USB port.
➢
Press the
PRINT SCREEN
softkey.
The output is started. A window to enter file name and path.
➢
Enter file name and path using the built-in auxiliary line editor or the external keyboard,
e.g.
F:\ display.wmf
and confirm the entry with the
ENTER
key.
➢
Press the
ENTER
key.
The screen content is stored on the memory stick under the specified file name.
During storing the softkeys are blanked. Operation in the menus can be continued when
the softkeys are displayed again.
The file can be used in other Windows applications, i.e. it can be linked.
gives an example of a stored display. In the example, two limit lines are active
and 4 markers positioned in the spectrum.
Summary of Contents for esu
Page 26: ...R S ESU Contents 6 Quick Start Guide 1302 6163 62 04 ...
Page 28: ......
Page 32: ...R S ESU 0 4 Quick Start Guide 1302 6163 62 04 ...
Page 88: ...R S ESU Preparing for Use Operating System Properties 2 38 Quick Start Guide 1302 6163 62 04 ...
Page 104: ...R S ESU Basic Operation Setting Parameters 4 12 Quick Start Guide 1302 6163 62 04 ...
Page 201: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix Appendix ...
Page 202: ...Quick Start Guide 1302 6163 62 04 R S ESU Appendix ...