Remote Control
R&S
®
ESR
992
User Manual 1175.7068.02 ─ 12
Manual operation:
See
SYSTem:RSW
<State>
Controls a repeated sweep of the E1 and MKPK HI/HL HP model commands (for
details on the commands refer to
Chapter 11.18, "GPIB Commands of HP Models
856xE, 8566A/B, 8568A/B and 8594E"
on page 1039). If the repeated sweep is OFF,
the marker is set without sweeping before.
This command is only available if a HP language is selected using
.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
SYSTem:RSW ON
Usage:
SCPI confirmed
Manual operation:
See
SYSTem:LANGuage
<Language>
This command defines the system language. Default is SCPI which defines the stan-
dard remote control behavior of the instrument. The other parameters are to simulate
the corresponding HP analyzer. The parameter "PSA89600" sets the device in a state
in which it can be used for IQ capturing from the 89600 vector signal analyzer soft-
ware.
Parameters:
<Language>
"SCPI" | "8560E" | "8561E" | "8562E" | "8563E" | "8564E" |
"8565E" | "8566A" | "8566B" | "8568A" | "8568A_DC" | "8568B" |
"8568B_DC" | "8591E" | "8594E" | "71100C" | "71200C" |
"71209A" | "PSA89600" | "PSA" | "ESA"
*RST:
SCPI
Example:
SYST:LANG '8560E'
Sets the system language to 8560E to simulate the HP analyzer.
Manual operation:
See
Note
: If you use
"PSA89600"
, you must switch to an HP language first before return-
ing to SCPI (in remote operation only). For the identical language "PSA", this inter-
mediate step is not necessary.
SYSTem:HPCoupling
<CouplingType>
Controls the default coupling ratios in the HP emulation mode for:
●
span and resolution bandwidth (Span/RBW) and
●
resolution bandwidth and video bandwidth (RBW/VBW)