Remote Control
R&S
®
ESR
866
User Manual 1175.7068.02 ─ 12
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:THR:STAT ON
Switches on the threshold line.
Manual operation:
See
See
DISPlay:MTABle
<DisplayMode>
This command turns the marker table on and off.
Parameters:
<DisplayMode>
ON
Marker table is displayed.
OFF
Marker table is not displayed.
AUTO
Marker table is only displayed if 2 or more markers are active.
*RST:
AUTO
Example:
To activate the table display:
DISP:MTAB ON
To query the current state of the marker table display:
DISP:MTAB?
Manual operation:
See
Using Delta Markers
Commands useful to control delta markers described elsewhere:
●
CALCulate<n>:DELTamarker<m>:AOFF
●
CALCulate<n>:DELTamarker<m>:LINK
●
CALCulate<n>:DELTamarker<m>:MAXimum:LEFT
●
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT
●
CALCulate<n>:DELTamarker<m>:MAXimum[:PEAK]
●
CALCulate<n>:DELTamarker<m>:MAXimum:RIGHt
●
CALCulate<n>:DELTamarker<m>:MINimum:LEFT
●
CALCulate<n>:DELTamarker<m>:MINimum:NEXT
●
CALCulate<n>:DELTamarker<m>:MINimum[:PEAK]
●
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt
●
CALCulate<n>:DELTamarker<m>:MODE
●
CALCulate<n>:DELTamarker<m>[:STATe]
●
CALCulate<n>:DELTamarker<m>:TRACe
●
Remote Commands in Spectrum Analyzer Mode