Receiver Mode
R&S
®
ESR
203
User Manual 1175.7068.02 ─ 12
"Ref. Marker"
Reference marker for delta markers. The marker values for the delta
marker are indicated relative to the specified reference marker.
The reference marker can either be another active marker, or a fixed
reference marker ("FXD", see
"Trace"
Trace for which the marker is to be set.
Remote command:
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
CALCulate<n>:DELTamarker<m>:TRACe
CALCulate<n>:DELTamarker<m>:MREF
All Marker Off ← Marker Wizard
Switches all markers off. It also switches off all functions and displays that are associ-
ated with the markers/delta markers.
Remote command:
All Marker Off
Switches all markers off. It also switches off all functions and displays that are associ-
ated with the markers/delta markers.
Remote command:
Marker Table
Defines how the marker information is displayed.
For more information, see
"On"
Displays the marker information in a table in a separate area beneath
the diagram.
"Off"
Displays the marker information within the diagram area.
"Aut"
(Default) The marker table is displayed automatically if more than 2
markers are active, and removed if only 1 or 2 markers are active.
This helps keep the information in the display clear.
Remote command:
Marker Info (On Off)
Turns the numerical marker information in the diagram area on and off.
Remote command:
DISPlay[:WINDow<n>]:MINFo:STATe
Tune to Marker
Defines the marker frequency as the new center frequency.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
Common Analysis Functions