Remote Control
R&S
®
ESR
701
User Manual 1175.7068.02 ─ 12
CALCulate<n>:DELTamarker<m>:MODE
<Mode>
This command selects the delta marker mode.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<Mode>
ABSolute
Delta marker position in absolute terms.
RELative
Delta marker position in relation to a reference marker.
*RST:
REL
Example:
CALC:DELT:MODE ABS
Absolute delta marker position.
CALCulate<n>:DELTamarker<m>[:STATe]
<State>
This command turns delta markers on and off.
If the corresponding marker was a normal marker, it is turned into a delta marker.
No suffix at DELTamarker turns on delta marker 1.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:DELT1 ON
Switches marker 1 to delta marker mode.
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
See
CALCulate<n>:DELTamarker<m>:TRACe
<TraceNumber>
This command selects the trace a delta marker is positioned on.
The corresponding trace must have a trace mode other than "Blank".
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Remote Commands in Receiver Mode