Remote Control
R&S
®
ESR
743
User Manual 1175.7068.02 ─ 12
CALCulate<n>:LIMit<k>:ACPower:ALTernate<channel>[:RELative]
<LowerLimit>,
<UpperLimit>
This command defines the limit for the alternate adjacent channels for adjacent chan-
nel power measurements. The reference value for the relative limit value is the mea-
sured channel power.
Note that the relative limit value has no effect on the limit check as soon as it is below
the absolute limit defined with
CALCulate<n>:LIMit<k>:ACPower:
. This mechanism allows automatic checking of
the absolute basic values of adjacent-channel power as defined in mobile radio stand-
ards.
Suffix:
<n>
.
Selects the measurement window.
<k>
irrelevant
<Channel>
1...11
the alternate channel
Parameters:
<LowerLimit>,
<UpperLimit>
first value: 0 to 100dB; limit for the lower and the upper alternate
adjacent channel
*RST:
0 DB
Example:
CALC:LIM:ACP:ALT2 30DB, 30DB
Sets the relative limit value for the power in the lower and upper
second alternate adjacent channel to 30 dB below the channel
power.
Manual operation:
See
CALCulate<n>:LIMit<k>:ACPower:ALTernate<Channel>[:RELative]:STATe
<State>
This command activates the limit check for the alternate adjacent channels for adjacent
channel power measurements. Before the command, the limit check must be activated
using
CALCulate<n>:LIMit<k>:ACPower[:STATe]
.
The result can be queried with
CALCulate<n>:LIMit<k>:ACPower:
. Note that a complete measurement must be
performed between switching on the limit check and the result query, since otherwise
no correct results are obtained.
Suffix:
<n>
.
Selects the measurement window.
<k>
irrelevant
<Channel>
1...11
the alternate channel
Remote Commands in Spectrum Analyzer Mode