Spectrum Measurements
R&S
®
ESR
293
User Manual 1175.7068.02 ─ 12
List Evaluation (On/Off) ← List Evaluation ← Spurious Emissions
Activates or deactivates the list evaluation.
Remote command:
CALCulate<n>:PEAKsearch|PSEarch:AUTO
Details On/Off ← List Evaluation ← Spurious Emissions
Configures the list contents.
On
Displays the whole list contents.
Off
Displays only the highest peaks (one peak per range).
Margin ← List Evaluation ← Spurious Emissions
Opens an edit dialog box to enter the margin used for the limit check/peak search.
Only peaks that exceed the limit and are larger than the specified margin are indicated
in the evaluation list.
Remote command:
CALCulate<n>:PEAKsearch|PSEarch:MARGin
Peaks per Range ← List Evaluation ← Spurious Emissions
Opens an edit dialog box to enter the number of peaks per range that are stored in the
list. Once the selected number of peaks has been reached, the peak search is stopped
in the current range and continued in the next range. The maximum value is 50.
Remote command:
CALCulate<n>:PEAKsearch|PSEarch:SUBRanges
Show Peaks ← List Evaluation ← Spurious Emissions
In the diagram, marks all peaks with blue squares that have been listed during an
active list evaluation.
Remote command:
CALCulate<n>:PEAKsearch|PSEarch:PSHow
Save Evaluation List ← List Evaluation ← Spurious Emissions
Opens the "ASCII File Export Name" dialog box and a submenu to save the result in
ASCII format to a specified file and directory.
Remote command:
ASCII File Export ← Save Evaluation List ← List Evaluation ← Spurious Emis-
sions
Saves the evaluation list in ASCII format to a specified file and directory.
Remote command:
Measurements