Maintenance, Storage and Disposal
R&S
®
ENV216
31
User Manual 5201.6693.14 ─ 04
In addition to the usual points, measurements should also be taken from the PE
contact of the mains plug to the bolt on the rear panel of the R&S
ENV216.
The resistance of the PE connection is calculated from the voltage reading divi-
ded by the supplied current.
6.5
Cleaning the Surface
Cleaning the outside surfaces of the device is best performed with a soft, lint-free
cloth.
Cleaning the Surface