![R&S CMW-KM5 Series User Manual Download Page 797](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802797.webp)
LTE PRACH Measurement
R&S
®
CMW-KM5xx/-KS5xx
797
User Manual 1173.9628.02 ─ 16
Example:
See
Configuring the Trigger System
Firmware/Software:
V2.0.10
Manual operation:
See
4.5.3.7
Limits (Modulation)
The following commands define limits for results which characterize the modulation
accuracy of the PRACH signal.
CONFigure:LTE:MEAS<i>:PRACh:LIMit:EVMagnitude
...................................................... 797
CONFigure:LTE:MEAS<i>:PRACh:LIMit:MERRor
..............................................................797
CONFigure:LTE:MEAS<i>:PRACh:LIMit:PERRor
.............................................................. 798
CONFigure:LTE:MEAS<i>:PRACh:LIMit:FERRor
.............................................................. 798
CONFigure:LTE:MEAS<i>:PRACh:LIMit:EVMagnitude
<RMS>, <Peak>
Defines upper limits for the RMS and peak values of the error vector magnitude (EVM).
Parameters:
<RMS>
Range:
0 % to 100 %
*RST:
17.5 %, ON
Default unit: %
Additional parameters: OFF | ON (disables the limit check | ena-
bles the limit check using the previous/default limit values)
<Peak>
Range:
0 % to 100 %
*RST:
35 %, OFF
Default unit: %
Additional parameters: OFF | ON (disables the limit check | ena-
bles the limit check using the previous/default limit values)
Example:
See
Firmware/Software:
V2.0.10
Manual operation:
See
CONFigure:LTE:MEAS<i>:PRACh:LIMit:MERRor
<RMS>, <Peak>
Defines upper limits for the RMS and peak values of the magnitude error.
Parameters:
<RMS>
Range:
0 % to 100 %
*RST:
17.5 %, OFF
Default unit: %
Additional parameters: OFF | ON (disables the limit check | ena-
bles the limit check using the previous/default limit values)
Command Reference