1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
394
User Manual 1173.9511.02 ─ 11
Return values:
<1_Reliabiltiy>
Conventions and General Information
. In list mode, a zero relia-
bility indicator indicates that the results in all measured seg-
ments are valid. A non-zero value indicates that an error occur-
red in at least one of the measured segments.
<2_SegReliability>
0 | 3 | 4 | 8
The segment reliability indicates whether one of the following
exceptions occurred in this segment:
0:
No error
3:
Signal overflow
4:
Signal low
8:
Synchronization error
If a combination of exceptions occurs, the most severe error is
indicated.
<3_RRI>, <4_Pilot>,
<5_ACKDSC>,
<6_AuxPilot>,
<7_DRC>, <8_Data>
RMS channel power values for the physical channels. Depend-
ing on the subtype 0/1 or 2 some channels exist or not. If not
available, no power measurement is possible therefore
NAV
is
returned.
Range:
-60 dB to +10 dB
Default unit: dB
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V2.0.20
Options:
R&S CMW-KM012
FETCh:EVDO:MEAS<i>:MEValuation:LIST:SEGMent<no>:CP:STATe?
Return the states of the channels for power measurement (CP).
The number to the left of each result parameter is provided for easy identification of the
parameter position within the result array.
Suffix:
<no>
.
1..200
The segment number must not exceed the total number of seg-
ments measured.
Return values:
<1_Reliabiltiy>
Conventions and General Information
. In list mode, a zero relia-
bility indicator indicates that the results in all measured seg-
ments are valid. A non-zero value indicates that an error occur-
red in at least one of the measured segments.
Command Reference
深圳德标仪器
135-1095-0799