1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
371
User Manual 1173.9511.02 ─ 11
Only the pilot part of a physical layer subtype 0/1 measurement is retrieved. For RRI
part and subtype 2 or 3 measurements see
.
Return values:
<Reliability>
<SigPilotLimit0> ...
<SigPilotLimit(SF-1)>
Return the exceeded limits as float values.
The number of results depends on the physical layer subtype
(see
): SF=15 for Subtype 0/1 and SF=31 for Subtypes 2 and 3 (only
INV values).
Range:
-70 dB to 0 dB
Default unit: dB
Usage:
Query only
Firmware/Software:
V2.0.10
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:CDE:ISIGnal:PILot:STATe?
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:CDE:QSIGnal:PILot:STATe?
Return the states of the code domain error (CDE) I-Signal and Q-Signal bar graphs.
Only the pilot part of a physical layer subtype 0/1 measurement is retrieved. For RRI
part and subtype 2 or 3 measurements see
.
Return values:
<Reliability>
<QSigPilotState0> ...
<QSigPilotStateSF>
NAV | ACTive | INACtive
The number of results depends on the physical layer subtype
(see
): SF=15 for Subtype 0/1 and SF=31 for Subtypes 2 and 3 (only
NAV values).
NAV
: No channel available
ACTive
: Active code channel
INACtive
: Inactive code channel
Usage:
Query only
Firmware/Software:
V2.0.10
3.5.4.16
CP Results
The following commands return the channel power (CP) results of the multi evaluation
measurement.
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:CP:STATe?
Return values:
<Reliability>