1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
323
User Manual 1173.9511.02 ─ 11
●
●
Parameters:
<ExpNomPower>
The range of the expected nominal power can be calculated as
follows:
Range (Expected Nominal Power) = Range (Input Power) +
External Attenuation - User Margin
Range:
-47 dBm to 55 dBm for the input power at RF COM
ports (please notice also the ranges quoted in the
data sheet)
*RST:
0 dBm
Default unit: dBm
Example:
See
Specifying General Measurement Settings
Firmware/Software:
V1.0.3.3
V3.0.10 max limits changed from +34 dBm to +55 dBm
Manual operation:
See
CONFigure:EVDO:MEAS<i>:RFSettings:UMARgin
<UserMargin>
Sets the margin that the R&S
CMW adds to the expected nominal power in order to
determine its reference power. The reference power minus the external input attenua-
tion must be within the power range of the selected input connector; refer to the data
sheet.
Parameters:
<UserMargin>
Range:
0 dB to (55 dB + External Attenuation - Expected
Nominal Power)
*RST:
0 dB
Default unit: dB
Example:
See
Specifying General Measurement Settings
Firmware/Software:
V1.0.3.3
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:PLSubtype
<PLSubtype>
Selects the physical layer protocol subtype.
This command is only relevant for the standalone scenario. For the combined signal
path scenario, use
Parameters:
<PLSubtype>
ST01 | ST2 | ST3
ST01:
subtype 0 or 1
ST2:
subtype 2
ST3:
subtype 3
*RST:
ST2
Command Reference