Remote Control
R&S
®
AFQ100A, R&S
®
AFQ100B
16
Supplement 1175.6978.02 ─ 04
c) Fill in the "Alias" name, the "VISA Resource String" and the "Device IP Address
or host name" as shown in the figure, and confirm with "OK".
The "Alias" name is assigned to the instrument.
d) Close the dialog.
The R&S
AFQ is now registered in the program and can be addressed via the
resource string or alias name.
9. In the main window, select "Connect".
R&S
VISA establishes the connection to the R&S
AFQ.
Now you can send settings to configure the instrument and receive its responses.
Note:
If the connection cannot be set up, R&S
VISA displays an error in the log
view.
Check whether the controller and the instrument are connected to the network (net-
work cable) and switched on. Correct spelling of the IP address or the computer
name can also be checked. For further error location, inform the network adminis-
trator. In large networks, specification of additional addresses may be required for
link setup, e.g. gateway and subnet mask, which are known to the network admin-
istrator.
For further information on the functions to read and write to an open session, as
well as the utility applications the software provides, see the R&S
VISA User Man-
ual.
Remote Control over LAN using VXI-11 Protocol