Remote control commands
R&S
®
NRPxxP
55
User Manual 1179.5760.02 ─ 02
TRIGger:SYNC:STATe
<state>
Available in:
●
Continuous average mode with
IMMediate
●
Trace, timeslot modes with
INTernal
Enable the synchronization state if you want to synchronize several power sensors
connected to a base unit. The synchronization works as follows.
The external trigger bus is blocked as long as a power sensor is in the "measuring"
state. The external trigger bus is enabled only when all power sensors are in the "wait-
ing for trigger" or "idle" state. Thus, the power sensors start to measure simultaneously.
Make sure that the measurement time is the same for all power sensors involved in the
measurement. Otherwise, the trigger bus is blocked by any power sensor that has
completed its measurement before the others and has returned to the "idle" state.
1
1 = measurement time
During the measurement, the trigger synchronization is high. For a continuous average
measurement, the measurement time is calculated as described in
"Calculating the measurement time"
For a trace or timeslot measurement, one measurement cycle is one sweep.
For details on trigger states, see
Chapter 8.5.2.1, "Trigger states"
Parameters:
<state>
*RST:
OFF
8.6
Configuring the measurement modes
In the following, the settings needed for configuring a measurement mode are descri-
bed.
Further information:
●
Chapter 8.7, "Configuring basic measurement parameters"
●
Chapter 8.5, "Controlling the measurement"
●
Chapter 8.9, "Querying measurement results"
8.6.1
Continuous average measurement
The continuous average mode measures the signal average power asynchronously
within definable time intervals (sampling windows). The aperture (width of the sampling
windows) can be defined.
Configuring the measurement modes