Remote control commands
R&S
®
NRPxxP
74
User Manual 1179.5760.02 ─ 02
8.7
Configuring basic measurement parameters
The following section describes the settings common for several measurement modes.
8.7.1
Configuring auto averaging
Describes the commands for automatic averaging in continuous average and burst
average measurements.
Remote commands:
................................................................................... 74
SENSe<Sensor>:AVERage:COUNt:AUTO
..........................................................................74
SENSe<Sensor>:AVERage:COUNt:AUTO:MTIMe
...............................................................75
SENSe<Sensor>:AVERage:COUNt:AUTO:NSRatio
.............................................................75
SENSe<Sensor>:AVERage:COUNt:AUTO:RESolution
........................................................ 75
SENSe<Sensor>:AVERage:COUNt:AUTO:SLOT
................................................................ 76
.................................................................................... 76
SENSe<Sensor>:AVERage:COUNt:AUTO:TYPE
................................................................ 76
SENSe<Sensor>:AVERage:TCONtrol
................................................................................ 77
...................................................................................77
SENSe<Sensor>:AVERage:COUNt
<count>
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Average count is often also called averaging factor, but it designates the same thing,
i.e the number of measured values that have to be averaged for forming the measure-
ment result.
Averaging is only effective, if
ON
is set.
Parameters:
<count>
Range:
1 to 1048576
*RST:
1024
Example:
AVER:COUN 1
SENSe<Sensor>:AVERage:COUNt:AUTO
<state>
Sets how the average count is determined.
Parameters:
<state>
ON
Auto averaging: the averaging factor is continuously determined
and set depending on the power level and other parameters.
OFF
Fixed filter: the previous, automatically determined averaging
factor is used.