Remote Control Commands
R&S
®
NRX
249
User Manual 1178.5566.02 ─ 07
Parameters:
<value>
*RST:
OFF
[SENSe<Sensor>:]TRACe:MEASurement:OFFSet:TIME
<value>
Sets the start time of the pulse analysis, referenced to delay set by
TRIGger<Measurement>[:CHANnel<Channel>]:DELay[:VALue]
.
By default, the parameters of the first detected pulse are determined. Using this com-
mand, you can determine parameters of another pulse. Make sure that the pulse
analysis does not begin until shortly before this pulse.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
Default unit: s
[SENSe<Sensor>:]TRACe:MID:OFFSet:TIME
<time>
Sets an offset to the start of the gate in which the pulse analysis is performed. See also
[SENSe<Sensor>:]TRACe:MID:TIME
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
0.0 to 30.0
*RST:
0.01
Default unit: s
[SENSe<Sensor>:]TRACe:MID:TIME
<time>
Sets the length of the gate in which the pulse analysis is performed. The gate start is
referenced to the trigger event.
Used to define a gate on a trace measurement, in which you want to perform a pulse
analysis. If the gate length equals the length of the trace measurement, the first detec-
ted pulse is analyzed. Using a gate, you can investigate successive pulses.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH