Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean>
ON (1)
Port <Ph_pt> is active in the respective measurement
OFF (0)
Port <Ph_pt> is inactive in the respective measurement
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT:M1L:DIFFmode
<Boolean>
CALCulate:FMODel:DELT<Ph_pt>:METHod M1L
) handles
single-ended and balanced measurement results differently. This command selects
between the two modes.
Parameters:
<Boolean>
OFF (0)
Single-ended mode
ON (1)
Differential mode
*RST:
OFF (0)
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT:MEASurement
<Measurement>
Defines whether single-ended or balanced ports are measured for Delta-L characteri-
zations.
Parameters:
<Measurement>
SINGleended | BALanced
SINGleended
Single-ended ports only
BALanced
Balanced ports only
*RST:
SINGleended
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT<Ph_pt>:METHod
<Method>
Selects the Delta-L+ characterization method to be used.
SCPI command reference