The effect of a reconnection depends on the selected
, the effect of a cable
movement depends on the selected
(on their respective models, to be precise).
For the METAS calibration, the VNA firmware reports the minimum possible number of
reconnections (depending on the selected calibration type) and no cable movements.
All other events of these types must be reported manually, using the "METAS Recon-
nection" dialog.
Add New Connection/Add Cable Movement
While a METAS calibration is active, these buttons notify the METAS VNA Tools about
additional connector reconnections and/or cable movements. The latter is only applica-
ble if cables were configured during the setup of the METAS calibration.
Remote command:
[SENSe<Ch>:]CORRection:METas:CABLe:REConnection
[SENSe<Ch>:]CORRection:METas:CABLe:MOVement
Measurement Filename
Allows you to configure the name of the measurement data file generated with the
measurement journal entry. For uniqueness, a timestamp is always appended.
5.12
Channel Config softtool
The "Channel Config" functions select, create and delete channels, configure the
source and receive ports of the R&S
ZNA, and optimize the measurement process.
Access:
Channel – [Channel
Config] key
5.12.1
Channels tab
Allows you to create and delete channels, to modify the channel state, and to select a
channel as the active channel.
You can monitor the channel activity using the
mand and the output signals at pins 8 to 11 of the User Port connector.
Background information
Refer to
Chapter 4.1.3.3, "Active and inactive traces and channels"
Channel Config softtool