Command reference
R&S
®
ZNA
1260
User Manual 1178.6462.02 ─ 20
To perform a METAS calibration, use the standard commands for manual calibration
(see
Chapter 7.3.14.10, "[SENSe:]CORRection..."
on page 1262). For the calibration
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
with METAS
calibration types
MTOSm | MUOSm | MOSM | MTRL
.
●
METAS calibration is not fully supported via remote control. The import of cal kits
with uncertainties must be performed from the analyzer GUI, and must be done
before the first METAS calibration.
[SENSe:]CORRection:METas:CABLe:CATalog?
Lists the available METAS cable models, identified by their file names (without
*.cable
extension) in the local METAS VNA Tools II database.
Usage:
Query only
Options:
R&S
ZNA-K50
Manual operation:
See
[SENSe<Ch>:]CORRection:METas:CABLe:MOVement
[<Ports>[, <Ports>]...]
While a METAS calibration is active for the related channel (see
on page 1261), this command notifies the METAS
VNA Tools II about additional cable movements at the specified ports.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Ports>
<integer>
First port number, where cables were moved
ALL
Cable movements at all calibrated ports
NONE
Clears the cable movement history in the METAS project
<Ports>
Additional port numbers, where cables were moved
Usage:
Setting only
Options:
R&S
ZNA-K50
Manual operation:
See
"Add New Connection/Add Cable Movement"
[SENSe<Ch>:]CORRection:METas:CABLe:REConnection
[<Ports>[, <Ports>]...]
While a METAS calibration is active for the related channel (see
on page 1261), this command notifies the METAS
VNA Tools II about additional connector reconnections at the specified ports.
SCPI command reference