GUI reference
R&S
®
ZNA
534
User Manual 1178.6462.02 ─ 20
"Detect Port Assignment"
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
"Default Port Assignment"
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:
DEFault
Temperature Compensation ← Cal. Settings
Available for inline calibration systems (ICS) only. Applies to all inline calibration units
(ICUs) that are connected to the ICS and that are used with factory characterization
data.
If checked, the R&S
ZNA firmware adjusts the factory characterization data according
to the ICU temperatures.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TEMPcomp
Characterization ← Cal. Settings
Displays all characterizations that are stored in the active cal unit. The "Factory" char-
acterization is available for all calibration units; it ensures an accurate calibration for all
standard applications. To account for modifications of the cal unit such as the connec-
tion of additional adapters, you can generate modified sets of characterization data
using the cal unit characterization wizard. See
Chapter 5.11.2.3, "Characterize Cal Unit
on page 567. By default, the R&S
ZNA uses the last generated cal unit char-
acterization.
For an inline calibration system, each connected inline calibration unit has its own set
of characterizations:
Tip:
If the characterization wizard is password-protected, the "Characterization" button
is unavailable. Use this functionality to prevent inadvertent activation of inappropriate
characterizations. See
See also
Chapter 4.5.5.3, "Characterization of calibration units"
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
SYSTem:COMMunicate:RDEVice:AKAL:CKIT
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
Query further cal unit properties:
SYSTem:COMMunicate:RDEVice:AKAL:DATE?
Cal softtool