Command reference
R&S
®
ZNA
1280
User Manual 1178.6462.02 ─ 20
Manual operation:
See
Test Port
Port Assignment 1
Port Assignment 2
2
Cal Unit Port 1
-
3
Cal Unit Port 2
auto-detected
4
-
auto-detected
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:TPORt
<TestPort1>[, <TestPort2>]...
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:
, but initially defines the port assignment only by its underlying test port set; the
connected calibration unit ports are
auto-detected
at the start of the corresponding
calibration sweep.
Suffix:
<Ch>
.
Number of the channel
<Asg>
Number of the port assignment
Parameters:
<TestPort1>
First test port number
<TestPort2>
Second test port number ...
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
<CalType>,
<Characterization>[, <CalName>]
Selects a calibration type and a cal unit characterization (cal kit file) for an automatic
calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<CalType>
FNPort | TOSM | FOPort | OPTPort | FRTRans | FTRans |
RTRans | REFL | RSHort | PFNPort | VUOSm | VPUosm |
CPUosm
Calibration Type
FNPort
Full n-port (UOSM) calibration
SCPI command reference