GUI reference
R&S
®
ZNA
568
User Manual 1178.6462.02 ─ 20
Controls in the Characterize Cal Unit dialog
Calibration Unit
Displays the connected calibration units. The R&S
ZNA auto-detects all calibration
units which are connected to one of its USB ports. If several cal units are connected,
one of them must be selected for characterization (active cal unit).
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Authentication
Allows you to set a password to protect the characterization dialog and the
from unauthorized access and operation. "Set Password..." opens a dia-
log to enter the password and activate password protection at the next time the "Set
Password" dialog is opened. Enter an empty string (no password) to deactivate pass-
word protection.
Tip:
A password also blocks a switchover of the active characterization during calibra-
tion; see
Remote command:
[SENSe:]CORRection:COLLect:AUTO:CKIT:PASSword
Characterization Data
Displays all characterizations which are stored on the active cal unit. The "Factory"
characterization is available for all calibration units; it ensures an accurate calibration
for all standard applications. Characterizations stored on an SD card (inserted at the
cal unit) are prefixed with "SD:".
Tip:
Characterizations stored on an SD card (inserted at the cal unit) are prefixed with
"SD:".
Cal softtool