GUI reference
R&S
®
ZNA
434
User Manual 1178.6462.02 ─ 20
Global Check
Activates or deactivates the global limit check including upper/lower limits and ripple
limits. The global limit check is a composite limit check over all traces of the current
recall set. The result of the global check appears in a popup box whenever "Global
Check" is selected.
●
"PASS" represents pass for all traces with enabled limit check. A trace without limit
lines or with disabled individual limit check always passes the global check.
●
"FAIL" means that the limit check for one or more traces failed.
Remote command:
TTL1 Pass / TTL2 Pass
Assigns the active trace to the low-voltage (3.3 V) TTL output signals at the User Port
(see
Monitoring a single trace
If "TTL1 Pass" ("TTL2 Pass") is selected and the trace passes all active limit checks,
then the TTL signal is applied to pin 13 (pin 14) of the User Port. If one of the limit
checks fails, then no TTL signal is generated.
Monitoring several traces
If a channel contains several traces, is possible to assign each of them to any TTL out-
put. The assignment divides the traces of the channel into four groups:
●
not assigned to signal 1 or signal 2
●
assigned to signal 1, but not to signal 2
●
assigned to signal 2, but not to signal 1
●
assigned to both signals
If several traces are assigned to a pass/fail signal, then the TTL signal is only gener-
ated if all traces are within their respective limits. It is switched off if one trace exceeds
those limits.
Application: Graduated quality check
The two pass/fail signals can be used to distinguish three quality levels of a DUT. The
test is performed on two identical traces Trc1 and Trc2 within the same channel. Trc1
is configured with a tighter, Trc2 with a looser set of limit lines. For Trc1 "TTL1 Pass" is
enabled, for Trc2 "TTL2 Pass".
●
TTL1: signal
If Trc1 passes (and so does Trc2), the quality of the DUT is good.
●
TTL1: no signal, TTL2: signal
If Trc1 fails but Trc2 passes, the quality of the DUT is still sufficient.
●
TTL1: no signal, TTL2: no signal
Lines softtool