Command reference
R&S
®
ZNA
1136
User Manual 1178.6462.02 ─ 20
Setting parameters:
<Filename>
String parameter containing the file name. If no path is specified,
the file is stored to the directory
C:\Users\Public\Documents\ZNA\Report
; the extension
*.zip is appended automatically.
*RST:
n/a
Example:
DIAG:DEV:STAT 'report_16032011_1120'
Generate a report and store it to
C:\Users\Public\Documents\ZNA\Report\
report_16032011_1120
. Use the
MMEMory...
commands to
rename, move, or delete the file.
Usage:
Setting only
Manual operation:
See
DIAGnostic:DUMP:SIZE
<DumpSize>
Determines the level of detail and hence the size of the dump files created in case of
firmware exceptions.
Parameters:
<DumpSize>
NONE | MINI | NORMal | LARGe | FULL
Either disables dump file creation (NONE) or determines the
level of detail.
Manual operation:
See
DIAGnostic:PRODuct:OPTion:INFO?
<Option>, <Detail>
Queries a property of an installed software option, identified by its name.
Query parameters:
<Option>
Option name, e.g.
'ZNA-K2'
.
<Detail>
DESCription | TYPE | ACTivation | EXPiration | KEY
You can query for an option's DESCription, its KEY and key
TYPE, and its ACTivation and EXPiration date (if applicable).
Example:
DIAGnostic:PRODuct:OPTion:INFO? 'ZNA-K2',DESC
On a R&S
ZNA this returns
'Time Domain Analysis'
(if
installed).
Usage:
Query only
Manual operation:
See
DIAGnostic:SERVice:RFPower
<Boolean>
Turns the internal source power at all ports and the power of all external generators on
or off. This command is equivalent to
.
SCPI command reference