Command reference
R&S
®
ZNA
1056
User Manual 1178.6462.02 ─ 20
Example:
Suppose
CALCulate1:TRANsform:DTFault:PEAK:COUNt?
returns 3, then
CALCulate1:TRANsform:DTFault:PEAK:DATA1; DATA2;
DATA3
returns the coordinates of the DtF limit violation peaks.
Options:
R&S
ZNA-K2
Manual operation:
See
CALCulate<Chn>:TRANsform:DTFault:PEAK:STATe
<DtfPeakState>
If the active trace of channel <Chn> is a Distance to Fault (DtF) trace, this command
allows to enable/disable DtF limit checking.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<DtfPeakState>
OFF|ON (0|1)
Example:
*RST; :CALCulate1:TRANsform:DTFault:STATe ON
makes the active trace of channel 1 a DtF trace.
CALCulate1:TRANsform:DTFault:PEAK:STATe ON
enables DtF limit checking.
Options:
R&S
ZNA-K2
Manual operation:
See
CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold
<PeakThreshold>
If the active trace of channel <Chn> is a Distance to Fault (DtF) trace, this command
allows to set the fault limit for DtF limit checking.
Use
CALCulate<Chn>:TRANsform:DTFault:PEAK:STATe
to enable DtF limit
checking.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<PeakThreshold>
Fault limit
*RST:
-20
dB
Default unit: dB
Example:
*RST; :CALCulate1:TRANsform:DTFault:STATe ON
makes the active trace of channel 1 a DtF trace.
CALCulate1:TRANsform:DTFault:PEAK:THReshold -30
sets the DtF limit to -30 dB.
Options:
R&S
ZNA-K2
Manual operation:
See
SCPI command reference