Command reference
R&S
®
ZNB/ZNBT
960
User Manual 1173.9163.02 ─ 62
CALCulate<Chn>:MARKer<Mk>:THReshold:STATe
<Active>
CALCulate<Chn>:MARKer<Mk>:REFerence:THReshold:STATe
<Boolean>
Activates or deactivates the marker-specific threshold for (single) peak searches with
the related marker (see
Use
CALCulate<Chn>:MARKer<Mk>:THReshold
or
MARKer<Mk>:REFerence:THReshold
to set the threshold value.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number (ignored for reference markers)
Parameters:
<Active>
Boolean
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:TYPE
<Mode>
CALCulate<Chn>:MARKer<Mk>:REFerence:TYPE
<Mode>
Sets the marker mode for the related marker (see
The marker must be created before using
CALCulate<Chn>:MARKer<Mk>[:STATe]
or
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number (ignored for reference markers)
Parameters:
<Mode>
NORMal | FIXed | ARBitrary
NORMal:
MARKer<Mk>:REFerence:SEARch:TRACking ON
er's stimulus value (
or
CALCulate<Chn>:MARKer<Mk>:REFerence:X?
automatically with every sweep, otherwise it is fixed. The marker
position (
CALCulate<Chn>:MARKer<Mk>:REFerence:Y?
to the corresponding response value, i.e. the marker is always
positioned on the trace.
The marker's stimulus value can be set using
CALCulate<Chn>:MARKer<Mk>:X
or
CALCulate<Chn>:MARKer<Mk>:REFerence:X
; the marker
automatically follows the trace.
FIXed:
freezes the marker at the position determined by the cur-
rent stimulus and response value. The response value stored
with the marker is not adjusted to subsequent sweeps. Tracking
is disabled. Stimulus and response value are stored with the
marker; they are not adjusted to subsequent sweeps and trace
format changes.
SCPI command reference