Command reference
R&S
®
ZNB/ZNBT
909
User Manual 1173.9163.02 ─ 62
RESonant
The 2x thru test coupon will be split and used directly for deem-
bedding (parameter value
3
).
This option may be more accurate when the fixture and 2x Thru
have the same impedance at every location
Manual operation:
See
"Test Coupons > Insertion Loss"
CALCulate:FMODel:ISD:COUPon:MEASure
For a fixture modeling with the ISD tool and coupon types SYMMetric2x | OPEN1x |
SHORt1x (see
CALCulate:FMODel:ISD:COUPon:TYPE
), this command starts the
measurement of the coupon at the active ports (see
).
Example:
See
Chapter 8.2.8, "Fixture modeling"
Usage:
Event
Manual operation:
See
"Measure / Measure Open / Measure Short"
CALCulate:FMODel:ISD:COUPon:MEASure:FILename
<String>
For a fixture modeling with the ISD tool and coupon types SYMMetric2x | OPEN1x |
SHORt1x (see
CALCulate:FMODel:ISD:COUPon:TYPE
), this command loads the
coupon properties from a Touchstone file.
Parameters:
<String>
Path to the Touchstone file, either absolute or relative to the
selected working directory (
Manual operation:
See
"Load File / 1x Open Preset / 1x Short Preset"
CALCulate:FMODel:ISD:COUPon:MEASure:FILename:CLEar
Invalidates the file path previously set using
CALCulate:FMODel:ISD:COUPon:MEASure:OPEN:
, or
CALCulate:FMODel:ISD:COUPon:MEASure:SHORt:FILename
.
Usage:
Event
Manual operation:
See
"Load File / 1x Open Preset / 1x Short Preset"
CALCulate:FMODel:ISD:COUPon:MEASure:OPEN
For a fixture modeling with the ISD tool and coupon type OPSHort1x (see
CALCulate:FMODel:ISD:COUPon:TYPE
), this command starts the measurement of
the
Open
coupon at the active ports (see
).
Usage:
Event
SCPI command reference