Command reference
R&S
®
ZNB/ZNBT
1107
User Manual 1173.9163.02 ─ 62
Parameters:
<Boolean>
0 – hide the "Info Window" (default)
1 – show the "Info Window"
Manual operation:
See
DISPlay:LAYout
<LayoutMode>
Arranges the diagrams in the screen, leaving the diagram contents unchanged.
Parameters:
<LayoutMode>
LINeup | STACk | HORizontal | VERTical | GRID
LINeup
–
the diagrams are arranged side by side.
STACk
–
the diagrams are arranged one on top of the other.
HORizontal
–
the diagrams are arranged in horizontal rows.
VERTical
–
the diagrams are arranged in vertical rows.
GRID
–
the diagrams are arranged as a rectangular matrix. The
number of rows and columns is as defined with command
Example:
See
Manual operation:
See
DISPlay:LAYout:APPLy
<LayoutId>
Selects a previously defined layout for display in the analyzer screen.
Parameters:
<LayoutId>
Integer value 1, 2 ...
Current number, as defined by
Example:
See
Manual operation:
See
"Additional Functionality: SCPI Commands"
DISPlay:LAYout:DEFine
<LayoutId>, <LayoutFormatMode>, <LayoutData>
DISPlay:LAYout:DEFine?
<LayoutId>
Creates a horizontal or vertical display layout and provides it with an identifier (<Layou-
tId>).
Layouts are defined row by row (horizontal layouts) or column by column (vertical lay-
outs).
●
A horizontal layout consists of N rows, each of height h
i
(i = 1 to N). The heights
are defined in units relative to the total height of the screen, i.e. their sum h
1
+ h
2
+ ... h
N
must be equal to 1.00.
Each row contains a selectable number of diagrams with independent widths w
ij
(j
= 1, 2 ...M(i)). The sum of the widths in each row must also match the screen width,
hence w
i1
+ w
i2
+ ... w
iM(i)
= 1.00 for all rows (i = 1 to N).
SCPI command reference