R&S ESU
Receiver Mode
4.58
Operating Manual 1302.6163.12 - 03
•
TDOMAIN SCAN (
only available with the R&S ESU-K53 option)
A selection between an algorithm for pulsed signal or for CW signals can be
made.
Remote command:
SWE:SPAC LIN
Start -
Entry of start frequency
The start frequency of a subrange must be equal to or greater than the stop fre-
quency of the previous subrange.
On entering the start frequency, the preceding scan range is – if necessary –
adapted automatically to avoid overlapping of scan ranges.
Remote command:
SCAN1:STAR 150 kHz
Stop -
Entry of stop frequency
The stop frequency of a subrange must be equal to or greater than the start fre-
quency of the subrange.
On entering the stop frequency, the preceding scan range is – if necessary –
adapted automatically to avoid overlapping of scan ranges.
Remote command:
SCAN1:STOP 30 MHz
Step Size -
Entry of step size
In the case of linear frequency increments, step widths between 1 Hz and the maxi-
mum R&S ESU frequency can be set. When a step size greater than the scan range
is entered (from start to stop), R&S ESU performs a measurement at the start and
stop frequency.
With logarithmic frequency increments, values between 0.1% and 100% can be set
with steps of 1/2/3/5.
With
STEP
AUTO
selected, the step size cannot be changed because it is automat-
ically set with respect to the IF bandwidth.
Remote command:
SCAN1:STEP 4 kHz
RES BW -
Entry of IF bandwidth
In the case of quasipeak, CISPR Average or CISPR RMS weighting, usually a fixed
bandwidth is set which cannot be changed (CISPR).
However, the coupling of the IF bandwidth to the frequency range can be cancelled
using softkey
CISPR RBW UNCOUPLED
in the
MEAS DETECTOR
menu.
Remote command:
SCAN1:BAND:RES 9 kHz
AUTO PULSE the FFT algorithm is suitable for all types off signals -
pulsed and continuous.
AUTO CW
the FFT algorithm is only suitable for continuous signals.
The measurement speed is higher than with AUTO
PULSE.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...